Kjeldahl method is of prime importance in our day to day life. It is one of the important parameters to be determined in industries like – Food, Feed, Milk / Dairy, Pharmaceutical, Soils and Fertilizers, Water / Effluent / Sludge treatment, Vaccines, Petroleum products, Rubber products etc.

TKN – Total Kjeldahl Nitrogen- is one of the primary methods and most widely used method of choice by millions even as on today. This wet chemistry method has been improved and now a day’s become far more efficient, safe, low cost and environmental friendly then it used to be in earlier days. Thanks to new technologies incorporated in latest instrumentation to make it effortless and more accurate, sensitive and productive method.

Kjeldahl method is primarily divided in to three steps – DIGESTION, DISTILLATION & TITRATION followed by result calculations and reporting. We provide complete solution for Total Kjeldahl Nitrogen (TKN) in all three steps; customer can select models depending upon their sample load, priorities and budget. They can select solutions for individual steps or for complete package as per their choice.

Digestion

Automatic Digestion - DKL Series

The DKL Series digesters are fully automatic units where manual operations have been drastically reduced: lowering and lifting of the samples takes place automatically DKL digesters work in accordance with a variety of Standards such as AOAC, ISO, EPA, DIN etc.

Fully Automatic

The auto-lift facilitates the operator by reducing strain and time spent in manual tasks. The customizable programs allow for repeatable processes to run fully unattended.

Safe And Robust

The separated control panel is protected from heat and spills ensuring reliability and a long lifespan. The innovative microprocessor controls the temperature of the block and is supported by a safety thermostat.

User Friendly

The intuitive control panel makes operating and monitoring every stage of the analysis effortless. Many standard methods are pre-loaded and more can be customized and saved.

Semi-automatic Digestion - DK Series

The DK Series consists of basic digestion blocks providing accuracy, high thermal homogeneity and minimum energy consumption. These digester are suitable for Kjeldahl digestion and selected models can be combined with accessories in order to perform COD and Trace Metal Determination applications.

Excellent Accuracy And Repeatability
  • The aluminum heating block ensures high thermal homogeneity heating up to 450°C.
  • Temperature selection with 1°C steps and a precision of ± 0.2°C.
  • Heating block temperature stability of ± 0.5°C.
  • Intuitive interface The bright digital display provides maximum visibility and easy reading of the information.
  • Supports the recording of up to 20 methods with 4 temperature ramps.
  • Easy to use interface with 4 buttons